Numeric values are assigned to observations that meet certain automated or human identified criterion. Positive flags are considered passing. The meaning of these identifiers are as follows:

Flag Description
General
2 Measurement data filled from backup data recording source
1 Data failed automatic QAQC but manually passed
0 Data passes all QAQC metrics
-1 Data manually removed
Calibration
-2 System flush
-3 Invalid identifier
-4 Drift between adjacent reference tank measurements exceeds limit
-5 Time elapsed between reference tank measurements exceeds limit
-6 Reference tank measurements out of range
2B 205
-10 O3_ppb measurement out of range (0, 250)
-11 Flow_CCmin flow rate out of range (1800, 3000)
-12 Cavity_T_C cell temperature out of range (0, 50)
-13 Time_UTC recent outage > 1 hour -- warmup period of 5 mins
GPS
20 position at vehicle storage
-21 Fix_Quality invalid gga fix
-22 N_Sat number of satellites < 4
-23 Status invalid rmc status
-24 Time_UTC recent outage > 5 mins -- warmup period of 1 min
Licor 6262 IRGA
-30 CO2d_ppm measurement out of range (0, 3000)
-31 Flow_mLmin flow rate out of range (395, 405)
-32 Cavity_T_C_IRGA cavity temperature out of range (0, 50)
-33 Cavity_P_kPa_IRGA cavity pressure out of range (50, 115)
-34 Time_UTC recent outage > 1 hour -- warmup period of 5 mins
Licor 7000 IRGA
-40 CO2d_ppm measurement out of range (0, 3000)
-41 Flow_mLmin flow rate out of range (395, 405)
-42 Cavity_T_C_IRGA cavity temperature out of range (0, 55)
-43 Cavity_P_kPa_IRGA cavity pressure out of range (50, 115)
-44 Time_UTC recent outage > 1 hour -- warmup period of 2 mins
Los Gatos Research NO2
-50 NO2_ppb measurement out of range (0, 1000)
-51 Cavity_P_torr cavity pressure out of range (296, 300)
-52 Cavity_T_C cavity temperature out of range (0, 50)
-53 Time_UTC recent out > 1 hour -- warmup period of 5 mins
Los Gatos Research UGGA
-60 CH4d_ppm measurement out of range (0, 1000)
-61 CO2d_ppm measurement out of range (0, 3000)
-62 H2O_ppm measurement out of range (0, 30000)
-63 Cavity_P_torr cavity pressure out of range (135, 145)
-64 Cavity_T_C cavity temperature out of range (5, 45)
-65 Time_UTC recent outage > 1 hour -- warmup period of 5 mins
Magee AE33
-70 BC6_ngm3 measurement out of range (0, 100000)
-71 Flow_Lmin flow rate out of range (4.9, 5.1)
-72 Time_UTC recent outage > 1 hour -- warmup period of 5 mins
MetOne ES642
-80 PM2.5_ugm3 measurement out of range (0, 100000)
-81 Flow_Lmin flow rate out of range (1.9, 2.1)
-82 Cavity_RH_pct cavity humidty out of range (0, 50)
-83 Status alarm status > 0
-84 Time_UTC recent outage > 1 hour -- warmup period of 5 mins
Teledyne T200
-90 NO_ppb measurement out of range (-3, 20000)
-91 NO2_ppb measurement out of range (-3, 20000)
-92 Flow_CCmin flow rate out of range (350, 600)
-93 O3_Flow_CCmin ozone flow rate out of range (50, 150)
-94 RCel_Pres_inHgA reaction cell pressure out of range (3, 6)
-95 Moly_T_C molybdenum converter temperature out of range (305, 325)
-96 PMT_T_C photomultiplier tube temperature out of range (5, 12)
-97 Box_T_C operating temperature out of range (5, 50)
-98 Time_UTC recent outage > 2 hours -- warmup period of 1 hour
Teledyne T300
-100 CO_ppb measurement out of range (0, 100000)
-101 Flow_CCmin flow rate out of range (500, 1000)
-102 Samp_Pres_inHgA sample pressure out of range (15, 35)
-103 Samp_T_C sample temperature out of range (10, 100)
-104 Bench_T_C optical bench temperature out of range (46, 50)
-105 Wheel_T_C filter wheel temperature out of range (66, 70)
-106 Box_T_C operating temperature out of range (5, 50)
-107 Time_UTC recent outage > 2 hours -- warmup period of 1 hour
Teledyne T400
-110 O3_ppb measurement out of range (0, 10000)
-111 Flow_CCmin flow rate out of range (500, 1000)
-112 Samp_Pres_inHgA sample pressure out of range (15, 35)
-113 Samp_T_C sample temperature out of range (10, 50)
-114 Box_T_C operating temperature out of range (5, 50)
-115 Time_UTC recent outage > 2 hours -- warmup period of 1 hour
Teledyne T500u
-120 NO2_ppb measurement out of range (0, 1000)
-121 Samp_Pres_inHgA sample pressure out of range (15, 35)
-122 Phase_T_C gas phase chemiluminescence chamber temperature out of range (15, 35)
-123 Box_T_C operating temperature out of range (5, 50)
-124 ARef_L_mm auto reference loss out of range (400, 1100)
-125 Time_UTC recent outage > 2 hours -- warmup period of 1 hour
TEOM 1400ab
-130 PM2.5_ugm3 measurement out of range (0, 300000)
-131 Time_UTC recent outage > 1 hour -- warmup period of 5 mins